ROLGMay 20, 2021

Localization and Control of Magnetic Suture Needles in Cluttered Surgical Site with Blood and Tissue

arXiv:2105.09481v1
Originality Incremental advance
AI Analysis

This addresses the need for minimal invasiveness and reduced side effects in surgical automation by enabling precise needle control in complex, realistic surgical settings.

The study tackled the problem of real-time visual localization and autonomous robotic control of magnetic suture needles in cluttered surgical environments with blood and tissue, achieving localization errors of 0.73 mm RMS in clean conditions and up to 2.72 mm RMS in challenging conditions, and tip tracking errors from 2.6 mm to 3.7 mm RMS for autonomous suture path following.

Real-time visual localization of needles is necessary for various surgical applications, including surgical automation and visual feedback. In this study we investigate localization and autonomous robotic control of needles in the context of our magneto-suturing system. Our system holds the potential for surgical manipulation with the benefit of minimal invasiveness and reduced patient side effects. However, the non-linear magnetic fields produce unintuitive forces and demand delicate position-based control that exceeds the capabilities of direct human manipulation. This makes automatic needle localization a necessity. Our localization method combines neural network-based segmentation and classical techniques, and we are able to consistently locate our needle with 0.73 mm RMS error in clean environments and 2.72 mm RMS error in challenging environments with blood and occlusion. The average localization RMS error is 2.16 mm for all environments we used in the experiments. We combine this localization method with our closed-loop feedback control system to demonstrate the further applicability of localization to autonomous control. Our needle is able to follow a running suture path in (1) no blood, no tissue; (2) heavy blood, no tissue; (3) no blood, with tissue; and (4) heavy blood, with tissue environments. The tip position tracking error ranges from 2.6 mm to 3.7 mm RMS, opening the door towards autonomous suturing tasks.

Foundations

The foundational work for this paper's niche, ranked by how specifically the neighbourhood builds on it — not by global fame.

Your Notes