LGApr 9, 2022

Motion Artifacts Correction from Single-Channel EEG and fNIRS Signals using Novel Wavelet Packet Decomposition in Combination with Canonical Correlation Analysis

arXiv:2204.04533v119 citationsh-index: 51
Originality Incremental advance
AI Analysis

This addresses motion artifact removal for wearable EEG and fNIRS sensors, which is an incremental improvement over existing methods.

The paper tackled motion artifact correction in single-channel EEG and fNIRS signals by proposing wavelet packet decomposition (WPD) and WPD combined with canonical correlation analysis (WPD-CCA), achieving improvements such as an average ΔSNR of 30.76 dB and η of 59.51% for EEG, and ΔSNR of 16.55 dB and η of 41.40% for fNIRS.

The electroencephalogram (EEG) and functional near-infrared spectroscopy (fNIRS) signals, highly non-stationary in nature, greatly suffers from motion artifacts while recorded using wearable sensors. This paper proposes two robust methods: i) Wavelet packet decomposition (WPD), and ii) WPD in combination with canonical correlation analysis (WPD-CCA), for motion artifact correction from single-channel EEG and fNIRS signals. The efficacy of these proposed techniques is tested using a benchmark dataset and the performance of the proposed methods is measured using two well-established performance matrices: i) Difference in the signal to noise ratio (ΔSNR) and ii) Percentage reduction in motion artifacts (η). The proposed WPD-based single-stage motion artifacts correction technique produces the highest average ΔSNR (29.44 dB) when db2 wavelet packet is incorporated whereas the greatest average η (53.48%) is obtained using db1 wavelet packet for all the available 23 EEG recordings. Our proposed two-stage motion artifacts correction technique i.e. the WPD-CCA method utilizing db1 wavelet packet has shown the best denoising performance producing an average ΔSNR and η values of 30.76 dB and 59.51%, respectively for all the EEG recordings. On the other hand, the two-stage motion artifacts removal technique i.e. WPD-CCA has produced the best average ΔSNR (16.55 dB, utilizing db1 wavelet packet) and largest average η (41.40%, using fk8 wavelet packet). The highest average ΔSNR and η using single-stage artifacts removal techniques (WPD) are found as 16.11 dB and 26.40%, respectively for all the fNIRS signals using fk4 wavelet packet. In both EEG and fNIRS modalities, the percentage reduction in motion artifacts increases by 11.28% and 56.82%, respectively when two-stage WPD-CCA techniques are employed.

Foundations

The foundational work for this paper's niche, ranked by how specifically the neighbourhood builds on it — not by global fame.

Your Notes